Z-2498-2018Class II

Allura Xper

Philips North America LLC / initiated 2018-06-11 / Terminated / root cause: software

Philips Healthcare began a recall of Allura Xper Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive... on . FDA classified it as Class II and gives the reason as "The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2498-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Allura Xper Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Medical Device Correction letter dated June 11, 2018 was sent to customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers, until a software revision that corrects this issue becomes available, users should verify that the desired shutter position is set when performing the run after the first time the procedure type is changed during a single examination. This can be accomplished by first selecting a different procedure and then reselect the original procedure on the Xper Module or on the Data Monitor. Customer shall ensure that all staff with access to the affected systems are informed of the contents of this Field Safety Notice. A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips. For questions contact your local Philips representative Technical Support Line 1-800-722-9377.
Quantity in commerce
459 affected systems
Distribution
US Nationwide Distribution
Product codes and lots
Allura 8.1.25 Allura 8.1.25.1 Allura 8.1.25.5 Allura 8.2.25 Allura 8.2.25.5 Allura 8.2.27

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2498-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2498-2018Class IITerminatedFDA Mandated
Z-2499-2018Class IITerminatedFDA Mandated
Z-2500-2018Class IITerminatedFDA Mandated