Z-2548-2018Class II

AperFix Femoral Implant Coring Removal Drill, labeled as the following: a. Femoral Removal Coring Reamer 9mm...

Cayenne Medical Inc. / initiated 2010-04-28 / Terminated / root cause: design

Cayenne Medical Inc. began a recall of AperFix Femoral Implant Coring Removal Drill, labeled as the following: a. Femoral Removal Coring Reamer 9mm, Model Number CM-7109; b. Femoral Removal Coring... on . FDA classified it as Class II and gives the reason as "A review of product complaints in 2010 identified a trend for the Coring Removal Drill of premature wear and/or breakage. This recall occurred in 2013". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2548-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cayenne Medical Inc.
Product
AperFix Femoral Implant Coring Removal Drill, labeled as the following: a. Femoral Removal Coring Reamer 9mm, Model Number CM-7109; b. Femoral Removal Coring Reamer 10mm, Model Number CM-7110; c. Femoral Removal Coring Reamer 11mm, Model Number CM-7111; d. Femoral Removal Coring Reamer 9 mm - Long, Model Number CM-7129; e. Femoral Removal Coring Reamer 10 mm - Long, Model Number CM-7130; f. Femoral Removal Coring Reamer 11 mm - Long, Model Number CM-7131; Orthopedic surgical instrument.
Product code
HTO Reamer
Reason for recall
A review of product complaints in 2010 identified a trend for the Coring Removal Drill of premature wear and/or breakage. This recall occurred in 2013.
Root cause tag
design
FDA root cause
Component design/selection
Action
Customers were contacted via email on approximately 10/10/2013 and were instructed to return any affected products on hand.
Quantity in commerce
1,444 units total
Distribution
Worldwide distribution. US nationwide, Denmark, Finland, Malaysia, Netherlands, Poland, Switzerland, and Turkey.
Product codes and lots
a. Model Number CM-7109, Lot Codes: 2091004, 3041001, 3151117, 3250806, 3291018, 4191101, 5181012, 6211011, 6270801, 7081012, 7131004, 7231005, 7281004, 7281019, 8161104, 8311131, 8311133, 8311134, 9081004, 9091102, 9141012, 9170902, 9220903, 9220910, 9271002, 10130801, 10140904, 10160901, 10211013, 11091001, 11161022, 11171022, 12030901, 12131001, 12281001, 12290910, 12291008, LNR;   b. Model Number CM-7110, Lot Codes: 2091005, 3151118, 3250807, 3291014, 4191102, 4201002, 5181002, 5181009, 6211013, 6270802, 7081013, 7131005, 7231001, 7281005, 7281020, 7311217, 9030903, 9080802, 9081005, 9091103, 9141013, 9161103, 9170903, 9220904, 9220911, 9271003, 10140905, 10160902, 10211001, 11071104, 11091002, 12030902, 12080802, 12131002, 12281002, 12291010, LNR;   c. Model Number CM-7111, Lot Codes: 2091006, 3151119, 3250808, 3291017, 4191103, 5181003, 5181013, 6211015, 7081014, 7131006, 7131007, 7140802, 7231006, 7281006, 7281021, 8161103, 8311128, 8311129, 8311130, 8311132, 9081006, 9091104, 9141014, 9220905, 9220912, 10140906, 10160804, 10160903, 10211011, 11091003, 12030903, 12131003, 12291009, LNR;   d. Model Number CM-7129, Lot Codes: 1130905, 4211103, 6211012, 7010909, 7081015, 7231010, 7281007, 7281022, 7311214, 8041105, 8060804, 8161102, 8311140, 8311141, 8311142, 9081007, 9091105, 9170904, 9220906, 10211004, 12131004, 12291011, LNR;   e. Model Number CM-7130, Lot Codes: 1130906, 3151120, 3291015, 5181010, 6211014, 7010910, 7081016, 7231007, 7281008, 7281023, 7311215, 8041106, 8060803, 8311135, 9081008, 9091106, 9170905, 9220907, 10211005, 12090903, 12131005, 12290912, 12291012, 90091106, LNR;    f. Model Number CM-7131, Lot Codes: 1130907, 2111207, 3151121, 3291016, 4211102, 5181011, 6211016, 7010911, 7081017, 7231008, 7281009, 7311216, 8041107, 8060805, 8161101, 8221103, 8311136, 8311137, 8311138, 8311139, 9081009, 9091107, 9170906, 9220908, 10211010, 11090904, 11161015, 11171015, 12171118, 12290911, 12291013, LNR;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2548-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2548-2018Class IITerminatedVoluntary: Firm initiated