ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000
Siemens Medical Solutions USA, Inc began a recall of ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000 on . FDA classified it as Class II and gives the reason as "During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2524-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Correction notification was distributed to customers on 5/30/18. The notification provided the following information to customers: Siemens will correct the error by inspecting all affected ARTIS Pheno systems. As an immediate measure to prevent movement stops during a C-arm rotation followed by a C-arm drop, the maximum rotation speeds will be reduced. This correction limits the maximum manual LAO/RAO rotation from 25/sec to 20/sec and Cran/Caud rotation from 20/sec to 10/sec. All other system parameters as well as 3D acquisition and Perivision/Peristepping remain unaffected. You may continue to use the ARTIS Pheno system while awaiting implementation of the corrective action. Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX004/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
- Quantity in commerce
- 40 in the United States
- Distribution
- The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.
- Product codes and lots
- 164033, 164071, 164034, 164095, 164057, 164054, 164055, 164080, 164072, 164088, 164048, 164094, 164068, 164019, 164013, 164073, 164025, 164026, 164027, 164060, 164083, 164021, 164066, 164075, 164074, 164084, 164086, 164098, 164028, 164053, 164063, 164064, 164024, 164049, 164018, 164022, 164061, 164100, 164097, 164011
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2524-2018 | Class II | Terminated | Voluntary: Firm initiated |