Z-2524-2018Class II

ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000

Siemens Medical Solutions USA, Inc. / initiated 2018-05-30 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000 on . FDA classified it as Class II and gives the reason as "During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2524-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Correction notification was distributed to customers on 5/30/18. The notification provided the following information to customers: Siemens will correct the error by inspecting all affected ARTIS Pheno systems. As an immediate measure to prevent movement stops during a C-arm rotation followed by a C-arm drop, the maximum rotation speeds will be reduced. This correction limits the maximum manual LAO/RAO rotation from 25/sec to 20/sec and Cran/Caud rotation from 20/sec to 10/sec. All other system parameters as well as 3D acquisition and Perivision/Peristepping remain unaffected. You may continue to use the ARTIS Pheno system while awaiting implementation of the corrective action. Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX004/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
Quantity in commerce
40 in the United States
Distribution
The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.
Product codes and lots
164033,  164071,  164034,  164095,  164057,  164054,  164055,  164080,  164072,  164088,  164048,  164094,  164068,  164019,  164013,  164073,  164025,  164026,  164027,  164060,  164083,  164021,  164066,  164075,  164074,  164084,  164086,  164098,  164028,  164053,  164063,  164064,  164024,  164049,  164018,  164022,  164061,  164100,  164097,  164011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2524-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2524-2018Class IITerminatedVoluntary: Firm initiated