Z-2571-2018Class II
Drill (Fluted-Hard Bone), for Quattro Link 2.9mm, labeled as: a. Quattro Spade Drill, for 2.9mm, Model Number...
Cayenne Medical Inc. / initiated 2013-10-10 / Terminated / root cause: design
Cayenne Medical Inc. began a recall of Drill (Fluted-Hard Bone), for Quattro Link 2.9mm, labeled as: a. Quattro Spade Drill, for 2.9mm, Model Number CM-9300; b. Quattro Drill 2.4mm, Hard Bone, Model... on . FDA classified it as Class II and gives the reason as "A review of complaints identified a trend for drills breaking". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2571-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cayenne Medical Inc.
- Product
- Drill (Fluted-Hard Bone), for Quattro Link 2.9mm, labeled as: a. Quattro Spade Drill, for 2.9mm, Model Number CM-9300; b. Quattro Drill 2.4mm, Hard Bone, Model Number CM-9324
- Product code
- HTW Bit, Drill
- Reason for recall
- A review of complaints identified a trend for drills breaking.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Customers were contacted on approximately 10/10/2013 and were instructed to return any affected products on hand.
- Quantity in commerce
- 229 devies total
- Distribution
- US distribution to AK, AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
- Product codes and lots
- a. Model Number CM-9300, Lot Number 2211211, 4271209, LNR, 7181408, 1221502, 4071521, 5091602, 7221613, 8301608, 9271623, 9271619, 9271621, 59068, 59069; b. Model Number CM-9324, Lot Number LNR, LRN, 11071203, 11281201, 4031303, 6061310, 9061302, 10081305, 1061401, 7181406, 9161409, 1221503, 1221501, 4071522, 6241501, 11051503, 3011650, 5091601, 7111613, 8301609, 9271618, 9271620, 11463, 11141
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2571-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.