Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code...
COVIDIEN MEDTRONIC began a recall of Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT on . FDA classified it as Class II and gives the reason as "Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2507-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT
- Product code
- GDW Staple, Implantable
- Reason for recall
- Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On May 22, 2018, all U.S. consignees received an UPDATED URGENT MEDICAL DEVICE RECALL and RETURN FORM notice via Federal Express or certified. The initial notice was sent to impacted customers on April 26, 2018. Customers are advised to take the following required actions: 1. Immediately discontinue use, and quarantine affected products listed 2. Return affected product as indicated. 3. Complete the Recalled Product Return Form even if you do not have inventory and submit by either method below: * Product purchased directly from Medtronic: [email protected] or fax to (800) 895-6140. *Product purchased through distributor: [email protected] or fax it to (203) 492-7719.
- Distribution
- Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta, Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom
- Product codes and lots
- N7L0852X, N7M0130X, N8A0145X, N8A0336X, N7L0931X, N7M0639X, N8A0796X & N8B0993X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2503-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2504-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2505-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2506-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2507-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2508-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2509-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2510-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2511-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2512-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2513-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2514-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2515-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2516-2018 | Class II | Ongoing | Voluntary: Firm initiated |