Z-2596-2018Class II

EZ1 DSP Virus Kit (48), IVD, REF 62724 intended for in vitro diagnostic use

Qiagen / initiated 2018-05-28 / Terminated

QIAGEN began a recall of EZ1 DSP Virus Kit (48), IVD, REF 62724 intended for in vitro diagnostic use on . FDA classified it as Class II and gives the reason as "Underfilled reagent well that could lead to incorrect results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2596-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Qiagen
Product
EZ1 DSP Virus Kit (48), IVD, REF 62724 intended for in vitro diagnostic use.
Product code
JJH Clinical Sample Concentrator
Reason for recall
Underfilled reagent well that could lead to incorrect results
FDA root cause
Under Investigation by firm
Action
On May 28, 2018 QIAGEN issued an Urgent Notice of Product Field Action and a Acknowledgement of Receipt form to customers. Actions to be taken by the customer/user/commercial partner are as follows: - If you have any remaining stocks, do not use them, contact our technical service for a free-of-charge replacement and dispose of the product in accordance with your national, local safety and environmental regulations. - Review the notice with your laboratory/medical director. - Forward the information to all individuals and departments within your organization. - If product was further distributed, forward a copy of the notice to the product end user. - QIAGEN also recommends a review of the corresponding results. Consider the possibility of under-quantification and false negative results and evaluate the need for retesting with different EZ1 DSP Virus Kit lots e.g. depending on the clinical presentation and history of the patient - Complete and return the Acknowledgement of Receipt form and email to: [email protected].
Quantity in commerce
287 pcs
Distribution
Domestic - MD Internationally - AE, AU , CZ, DE, DZ, ES, FR, GB, HK , IE, IT, NO, RO & SA
Product codes and lots
UDI - 04053228000099 Lot - 160020069 Expiration - 11/01/19

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2596-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2596-2018Class IITerminatedVoluntary: Firm initiated