Instrument trays used to store AltiVate Small Shell Trials (FA S ALTIVATE RSP HUM TRLS SM). Metal container...
Encore Medical, Lp began a recall of Instrument trays used to store AltiVate Small Shell Trials (FA S ALTIVATE RSP HUM TRLS SM). Metal container with lid used to organize and transport... on . FDA classified it as Class II and gives the reason as "A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide coating is not biocompatible". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2600-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Instrument trays used to store AltiVate Small Shell Trials (FA S ALTIVATE RSP HUM TRLS SM). Metal container with lid used to organize and transport instrumentation to be used in orthopedic surgery. Inner trays also used to facilitate cleaning and sterilization of the instrumentation by end users.
- Product code
- LRO General Surgery Tray
- Reason for recall
- A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide coating is not biocompatible.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, DJO LLC, sent an "URGENT FIELD SAFETY NOTICE" letter dated 6/26/2018 to its direct accounts by email on 06/26/2018. The letter described the product, problem and actions to be taken. The customers were instructed to return product; contact Customer Service as soon as possible to request an RMA for return and exchange of the full instrument set; and to coordinate the exchange of products with Customer Service. If you do not have these instrument sets or the number of sets identified above, you are required to complete the attached form and email it to [email protected]. ALL ACTIVITIES OUTLINED ABOVE SHOULD BE COMPLETED BY JULY 30, 2018. If you have any questions, please call (512) 834-6255 or by email to [email protected].
- Quantity in commerce
- 30 units
- Distribution
- US Distribution to states of; AZ, CA, CO, FL, IN, NJ, NY, NV, MI, MN, MO, OH, RI, SC, TN, TX, UT and WA and country of; Japan.
- Product codes and lots
- 804-99-129 Vendor Lot #22747003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2600-2018 | Class II | Terminated | Voluntary: Firm initiated |