Z-2816-2018Class II

6" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer

ICU MEDICAL de MEXICO, S. de R.L. de C.V. / initiated 2018-03-01 / Terminated

ICU Medical de Mexico, S.A. de C.V. began a recall of 6" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer on . FDA classified it as Class II and gives the reason as "ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2816-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ICU MEDICAL de MEXICO, S. de R.L. de C.V.
Product
6" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer.
Product code
FMG Stopcock, I.V. Set
Reason for recall
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.
FDA root cause
Other
Action
ICU Medical notified customers of the recall on about 03/01/2018 via URGENT: Medical Device Recall Notification. The notification informed customers that this was an update and expansion to previously initiated recall. Instructions included to inspect inventory for affected devices, arrange for the return of affected devices, and complete and return the Urgent Medical Device Recall Notification Response Form to [email protected].
Quantity in commerce
300 pouches
Distribution
US distribution to AZ, MA, NC, OH, TX, VA, WA. Worldwide distribution to Belgium, France, Germany.
Product codes and lots
Item Number AM6130.   a. Lot Number 3414534, UDI Number (01) 1 0887709 08771 9 (17) 220201 (30) 50 (10) 3414534, Expiration Date 02/01/2022;  b. Lot Number 3442461, UDI Number (01) 1 0887709 08771 9 (17) 220501 (30) 50 (10) 3442461, Expiration Date 05/01/2022;  b. Lot Number 3509517, UDI Number (01) 1 0887709 08771 9 (17) 220801 (30) 50 (10) 3509517, Expiration Date 08/01/2022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2816-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2797-2018Class IITerminatedVoluntary: Firm initiated
Z-2798-2018Class IITerminatedVoluntary: Firm initiated
Z-2799-2018Class IITerminatedVoluntary: Firm initiated
Z-2800-2018Class IITerminatedVoluntary: Firm initiated
Z-2801-2018Class IITerminatedVoluntary: Firm initiated
Z-2802-2018Class IITerminatedVoluntary: Firm initiated
Z-2803-2018Class IITerminatedVoluntary: Firm initiated
Z-2804-2018Class IITerminatedVoluntary: Firm initiated
Z-2805-2018Class IITerminatedVoluntary: Firm initiated
Z-2806-2018Class IITerminatedVoluntary: Firm initiated
Z-2807-2018Class IITerminatedVoluntary: Firm initiated
Z-2808-2018Class IITerminatedVoluntary: Firm initiated
Z-2809-2018Class IITerminatedVoluntary: Firm initiated
Z-2810-2018Class IITerminatedVoluntary: Firm initiated
Z-2811-2018Class IITerminatedVoluntary: Firm initiated
Z-2812-2018Class IITerminatedVoluntary: Firm initiated
Z-2813-2018Class IITerminatedVoluntary: Firm initiated
Z-2814-2018Class IITerminatedVoluntary: Firm initiated
Z-2815-2018Class IITerminatedVoluntary: Firm initiated
Z-2816-2018Class IITerminatedVoluntary: Firm initiated