Z-2535-2018Class II

HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system

Elekta Limited / initiated 2018-05-16 / Terminated

Elekta, Inc. began a recall of HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a... on . FDA classified it as Class II and gives the reason as "Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2535-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy
Product code
JAI Couch, Radiation Therapy, Powered
Reason for recall
Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning.
FDA root cause
Under Investigation by firm
Action
The letter from Elekta dated May 2018 was issued via email 5/16/2018 and was flagged "URGENT IMPORTANT FIELD SAFETY NOTIFICATION. The letter identified the affected product, problem and actions to be taken.
Quantity in commerce
725 devices
Distribution
Worldwide Distribution - US Distribution of the device was made nationwide. There was government distribution but no military distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Democratic People's Republic of Korea, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Sweden, Thailand, Turkey, United Kingdom, and Viet Nam.
Product codes and lots
UDI 04056719001698.  Serial numbers: 42380, 42384, 42626, 03005, 42009, 42012, 42013, 42062, 42068, 42106, 42107, 42126, 42153, 42217, 42224, 42292, 42293, 42301, 42435, 42491, 42496, 42517, 42615, 42620, 42690, 42691, 42731, 03019, 42092, 42222, 42356, 42382, 42601, 42649, 42658, 42743, 42289, 03000, 42311, 42586, 42471, 03017, 03022, 42057, 42058, 42129, 42159, 42187, 42279, 02982, 02986, 02991, 42004, 42113, 42178, 42189, 42202, 42207, 42209, 42212, 42213, 42216, 42220, 42416, 42417, 42418, 42419, 42420, 42463, 42652, 42704, 42706, 02987, 03008, 03018, 42000, 42008, 42032, 42048, 42064, 42070, 42074, 42084, 42102, 42115, 42118, 42134, 42137, 42142, 42145, 42146, 42152, 42174, 42177, 42237, 42312, 42329, 42330, 42341, 42363, 42369, 42371, 42375, 42383, 42394, 42409, 42446, 42450, 42453, 42457, 42464, 42474, 42501, 42509, 42519, 42523, 42524, 42550, 42553, 42587, 42611, 42612, 42643, 42647, 42662, 42674, 42681, 42694, 42721, 42722, 42069, 03029, 42036, 42709, 42710, 42139, 42162, 42591, 42592, 42421, 42669, 42011, 03006, 42007, 42109, 42173, 42193, 42282, 42392, 42405, 42427, 42428, 42504, 42535, 42602, 42603, 42644, 42680, 42685, 02998, 03003, 03033, 42015, 42053, 42053, 42065, 42085, 42091, 42094, 42105, 42124, 42198, 42219, 42271, 42334, 42386, 42388, 42397, 42449, 42507, 42574, 42618, 42635, 42687, 42692, 03024, 42121, 42338, 42377, 42412, 42545, 42546, 42627, 42444, 42498, 03020, 42027, 42061, 42075, 42079, 42214, 42248, 42286, 42305, 42314, 42413, 42455, 42481, 42482, 42502, 42505, 42518, 42520, 42522, 42527, 42561, 42564, 42571, 42577, 42584, 42598, 42609, 42630, 42637, 42638, 42670, 42699, 42600, 42604, 42682, 42192, 42381, 02960, 02961, 02962, 02995, 03030, 42002, 42021, 42025, 42033, 42035, 42089, 42127, 42151, 42155, 42200, 42204, 42221, 42239, 42247, 42281, 42325, 42326, 42360, 42361, 42365, 42366, 42424, 42454, 42476, 42477, 42478, 42484, 42485, 42494, 42499, 42526, 42531, 42555, 42575, 42678, 42724, 42733, 42734, 02993, 02997, 03002, 03004, 03031, 42001, 42016, 42028, 42029, 42037, 42042, 42047, 42049, 42050, 42051, 42054, 42055, 42063, 42077, 42080, 42083, 42087, 42101, 42103, 42117, 42140, 42147, 42150, 42156, 42157, 42163, 42165, 42166, 42169, 42179, 42180, 42182, 42183, 42186, 42195, 42196, 42197, 42223, 42227, 42231, 42234, 42235, 42236, 42240, 42243, 42245, 42249, 42250, 42251, 42253, 42254, 42256, 42260, 42261, 42262, 42264, 42265, 42266, 42267, 42268, 42269, 42270, 42272, 42276, 42291, 42298, 42300, 42304, 42331, 42333, 42336, 42337, 42347, 42349, 42350, 42353, 42354, 42355, 42358, 42362, 42378, 42379, 42385, 42389, 42398, 42399, 42400, 42401, 42402, 42403, 42404, 42406, 42436, 42439, 42440, 42441, 42442, 42443, 42458, 42459, 42460, 42472, 42483, 42493, 42511, 42512, 42536, 42537, 42538, 42539, 42554, 42557, 42562, 42565, 42567, 42568, 42569, 42570, 42580, 42581, 42582, 42597, 42610, 42660, 42663, 42672, 42696, 42697, 42698, 42701, 42708, 42226, 42018, 42135, 42185, 42252, 42273, 42284, 42364, 42367, 42374, 42488, 42618, 42648, 42677, 42525, 42719, 42081, 42090, 42093, 42095, 42110, 42114, 42133, 42181, 42332, 42671, 42576, 02580, 02977, 03016, 42116, 42396, 42532, 42689, 42010, 42073, 42295, 42296, 42661, 42753, 42176, 42164, 42588, 42589, 42590, 42599, 02963, 03028, 42373, 02976, 02979, 02988, 03015, 03021, 03027, 42022, 42034, 42038, 42041, 42088, 42104, 42348, 42352, 42372, 42452, 42468, 42238, 42346, 42302, 42547, 42560, 42412, 42490, 42432, 42433, 42659, 42006, 42024, 42060, 42082, 42194, 42232, 42263, 42318, 42342, 42344, 42351, 42368, 42395, 42534, 42628, 42642, 42673, 42688, 42693, 42702, 42199, 42573, 42684, 42170, 02985, 42078, 42138, 42158, 42161, 42188, 42215, 42242, 42275, 42308, 42327, 42370, 42414, 42425, 42445, 42470, 03032, 42120, 42313, 42533, 42641, 42056, 42306, 42530, 42665, 42667, 42707, 02964, 02969, 02972, 02984, 42290, 42495, 42563, 42607, 42616, 42629, 42668, 42675, 42725, 42739, 02966, 02968, 02970, 02971, 02973, 02974, 02981, 02990, 02992, 02996, 03001, 03007, 03012, 03013, 03014, 03025, 03026, 42014, 42017, 42020, 42023, 42030, 42040, 42043, 42044, 42045, 42046, 42052, 42066, 42067, 42071, 42072, 42096, 42097, 42099, 42100, 42111, 42112, 42122, 42123, 42125, 42130, 42131, 42136, 42141, 42143, 42144, 42148, 42149, 42154, 42160, 42171, 42172, 42175, 42190, 42191, 42203, 42205, 42206, 42208, 42211, 42218, 42225, 42228, 42229, 42230, 42233, 42241, 42244, 42255, 42257, 42258, 42274, 42277, 42278, 42280, 42285, 42287, 42288, 42294, 42297, 42299, 42303, 42307, 42309, 42310, 42315, 42320, 42322, 42323, 42324, 42328, 42335, 42339, 42340, 42345, 42357, 42359, 42390, 42391, 42393, 42407, 42408, 42411, 42423, 42426, 42429, 42431, 42437, 42438, 42448, 42456, 42467, 42469, 42473, 42475, 42479, 42486, 42486, 42487, 42489, 42492, 42497, 42500, 42506, 42510, 42515, 42528, 42529, 42540, 42541, 42542, 42543, 42544, 42548, 42552, 42556, 42558, 42559, 42566, 42572, 42578, 42579, 42583, 42585, 42605, 42606, 42613, 42614, 42619, 42621, 42632, 42633, 42634, 42639, 42640, 42651, 42653, 42654, 42655, 42656, 42657, 42686, 42700, 42703, 42714, 42717, 42720, 42727, 42128, 42593, 42617, and 42625.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2535-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2535-2018Class IITerminatedVoluntary: Firm initiated