Z-2575-2018Class II

Liquid Cardiac Control - Level 3, cat. no. CQ5053

Randox Laboratories, Limited / initiated 2018-06-08 / Terminated

Randox Laboratories Ltd. began a recall of Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac... on . FDA classified it as Class II and gives the reason as "The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2575-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.
FDA root cause
Under Investigation by firm
Action
On June 8, 2018, the firm issued an Urgent Medical Device Recall letter to affected consignees. The letter stated that Randox has confirmed Liquid Cardiac Control CQ5053 Lot: 4245CK is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. Customers were instructed to do the following: " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Update kits with corrected IFU's excluding Troponin T values. " Complete and return the vigilance response section of this form to [email protected] within five working days.) In addition, distributors were instructed to send a copy of the letter to all affected customers and to those who need to be aware within their organization.
Quantity in commerce
1 kit (US), 602 kits (OUS)
Distribution
Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam
Product codes and lots
Lot 4245CK.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2575-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2575-2018Class IITerminatedVoluntary: Firm initiated