Z-2525-2018Class II

AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed...

Siemens Medical Solutions USA, Inc. / initiated 2018-05-25 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and... on . FDA classified it as Class II and gives the reason as "Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2525-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens issued on May 25, 2018 an URGENT: MEDICAL DEVICE CORRECTION customer notification to affected customers and advised a hardware update will be initiated . Users advised of the problem, potential risks and action to be taken. Siemens will correct the error by replacing all affected cooling units with Update Instruction AX076/17/S. Users may continue to use the Artis system while awaiting implementation of the corrective action.
Quantity in commerce
6 units US
Distribution
HI , IL, MI, NC, NY, VA, WI
Product codes and lots
Laird Cooling Unit for SSFD: Material # 7555118.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2525-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2525-2018Class IITerminatedVoluntary: Firm initiated
Z-2526-2018Class IITerminatedVoluntary: Firm initiated