Z-2584-2018Class II
Revolution EVO 3.6, labeled as a. HVY WAUK; b. LONG HINO; c. LONG WAUK; d. MID BJG; e. MID WAUK
GE Healthcare / initiated 2018-05-24 / Terminated
GE Healthcare, LLC began a recall of Revolution EVO 3.6, labeled as a. HVY WAUK; b. LONG HINO; c. LONG WAUK; d. MID BJG; e. MID WAUK on . FDA classified it as Class II and gives the reason as "Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2584-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Revolution EVO 3.6, labeled as a. HVY WAUK; b. LONG HINO; c. LONG WAUK; d. MID BJG; e. MID WAUK.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
- FDA root cause
- Component change control
- Action
- GE Healthcare notified customers on about 05/24/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter outlined the reason for recall and notified customers that a GE Healthcare representative would be contacting them to arrange for the correction of the devices. Any questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or a local Service Representative.
- Quantity in commerce
- 42 units total
- Distribution
- Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
- Product codes and lots
- a. HVY WAUK, Serial Number CJRBX1700130CN (System ID 770836EVO, UDI 01008406821097961117120021CJRBX1700130CN), Serial Number CJRBX1800005CN (System ID 918635REVO, UDI 01008406821097961118010021CJRBX1800005CN); b. LONG HINO, Serial Number RE36A1800042YC (System ID 864560REV1, UDI 01008406821097961118020021RE36A1800042YC), Serial Number RE36A1800036YC (System ID 216448REVO, UDI 01008406821097961118020021RE36A1800036YC); c. LONG WAUK, Serial Number CJRBX1800037CN (System ID 740383REVO, UDI 01008406821097961118040021CJRBX1800037CN), Serial Number CJRBX1800022CN (System ID 317415REVO, UDI 01008406821097961118020021CJRBX1800022CN), Serial Number CJRBX1800013CN (System ID 423844REVO, UDI 01008406821097961118010021CJRBX1800013CN), Serial Number CJRBX1700132CN (System ID 719557REVO, UDI 01008406821097961117120021CJRBX1700132CN), Serial Number CJRBX1800008CN (System ID 510248EVOCT, UDI 01008406821097961118010021CJRBX1800008CN); d. MID BJG, Serial Number 500918HM8 (System ID 918623REVO, UDI Not Available), Serial Number CBCGG1700116HM (System ID 704834EVO, UDI 01008406821097961117120021CBCGG1700116HM), Serial Number CJRBX1800022CN (System ID 864560REVOL, UDI 01008406821097961118020021CJRBX1800022CN), Serial Number CBCGG1700097HM (System ID 979BRENCT1, UDI 01008406821097961117110021CBCGG1700097HM), Serial Number CBCGG1700094HM (System ID 405779REVO, UDI 01008406821097961117110021CBCGG1700094HM), Serial Number CBCGG1700107HM (System ID 217283HREVO, UDI 01008406821097961117110021CBCGG1700107HM), Serial Number CBCGG1700102HM (System ID 303SWDEVO, UDI 01008406821097961117110021CBCGG1700102HM), Serial Number CBCGG1700112HM (System ID 908788EVO, UDI 01008406821097961117120021CBCGG1700112HM), Serial Number 490670HM7 (System ID 718226REVO, UDI Not Available); e. MID WAUK, Serial Number CJRBX1800017CN (System ID 478454EVO, UDI 01008406821097961118020021CJRBX1800017CN), Serial Number CJRBX1800028CN (System ID 210MASHREVO, UDI 01008406821097961118030021CJRBX1800028CN), Serial Number CJRBX1800014CN (System ID 601968REVO, UDI 01008406821097961118010021CJRBX1800014CN).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2578-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2579-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2580-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2581-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2582-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2583-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2584-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2585-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2586-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2587-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
