Z-2662-2018Class II

Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism

Abbott Medical / initiated 2018-04-13 / Terminated

Smiths Medical ASD Inc. began a recall of Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism on . FDA classified it as Class II and gives the reason as "Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2662-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism
Product code
FRN Pump, Infusion
Reason for recall
Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.
FDA root cause
Process change control
Action
Customers were notified of the recall on about 04/13/2018 via URGENT MEDICAL DEVICE RECALL NOTICE letter. Instructions included to evaluate current inventory for affected Barrel Clamp Guide, review and complete the Recall Response Form, return affected Barrel Clamp Guides to Stericycle for destruction, and if the affected Barrel Clamp Guides have already been installed in pumps, continue to use the pumps, but utilize the Syringe Verification Reference Tool provided with the Recall Notice until the pumps are repaired. Customers were also instructed to notify customers if the affected devices were further distributed. Questions regarding the recall can be directed to Stericycle at [email protected].
Quantity in commerce
276 units
Distribution
US distribution to AL, AZ, CA, DC, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA. International distribution to Canada and Lebanon.
Product codes and lots
Barrel Clamp Guide mechanism number: G6000716;   Barrel Clamp Guide lot numbers: P0239588, P0321444, and P0462421

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2662-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2662-2018Class IITerminatedVoluntary: Firm initiated