Z-2590-2018Class II

Final Pack, Esteem II, Sound Processor Model 2001, containing 2 Sterile Wipes 002866-001 (which expired in...

Envoy Medical Corporation / initiated 2018-06-28 / Terminated / root cause: packaging

Envoy Medical Corporation began a recall of Final Pack, Esteem II, Sound Processor Model 2001, containing 2 Sterile Wipes 002866-001 (which expired in April 2018). An Implantable Middle Ear Hearing Device on . FDA classified it as Class II and gives the reason as "Sterile Wipes packed in these four Esteem II Sound Processors had an earlier expiration date than the final pack and the Esteem II Sound Processors. The Sterile Wipes expired prior to the date labeled on the final pack". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2590-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Envoy Medical Corporation
Product
Final Pack, Esteem II, Sound Processor Model 2001, containing 2 Sterile Wipes 002866-001 (which expired in April 2018). An Implantable Middle Ear Hearing Device.
Product code
OAF Implant, Hearing, Active, Middle Ear, Totally Implanted
Reason for recall
Sterile Wipes packed in these four Esteem II Sound Processors had an earlier expiration date than the final pack and the Esteem II Sound Processors. The Sterile Wipes expired prior to the date labeled on the final pack.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
The firm notified its customers via email on 06/18/2018 and 06/19/2018. The 06/18/2018 email to a customer stated Envoy Medical would like to have the devices the customer has on site returned, as apparently, it was confirmed that each device is packaged with a wipe that shows expired just on the wipe only. Envoy will replace the recalled devices by overnight shipping replacements. The 06/19/2018 email to a customer stated it was discovered that the sound processors (primary and backup) were previously sent to the site have a wipe in them that show an expired date on the packaging. The email also stated this is something that doesn't affect the actual sound processor itself as those are good, but just to be on the safe side, make sure the 2 sound processors currently at the site (serial #s 213658 and 213729) are returned to Envoy Medical ASAP so we can have them repackaged with wipes within their expiration dates added to the sound processor packaging. Envoy will send out 2 new Sound Processors (primary and backup). Both emails contained Envoy Medical contact information for the consignees in case any questions or concerns arose.
Quantity in commerce
4 Units
Distribution
US Distribution to states of: FL, OR
Product codes and lots
Lot Numbers: EMC0006145, EMC0006155, & EMC0006217

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2590-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2590-2018Class IITerminatedVoluntary: Firm initiated