Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000...
Boston Scientific Corporation began a recall of Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0 on . FDA classified it as Class II and gives the reason as "Reports of unintended continuation of radiofrequency (RF) energy delivery from the Cardiac ablation system after the foot switch has been released". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0202-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,
- Product code
- OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter
- Reason for recall
- Reports of unintended continuation of radiofrequency (RF) energy delivery from the Cardiac ablation system after the foot switch has been released.
- Action
- On 06/20/2018, a system notice was sent to customers, informing them that releasing the foot switch may not stop RF delivery. If RF delivery does not stop as intended, press and release the RF Power Control button. Customers were advised to report occurrence and clinical outcome of unintended RF delivery to the firm:United States (1-800-949-6708), Europe, Middle East, and Africa (+800-5555-7707), Asia Pacific (+65-6418-8878). Notification outside of the U.S. started 06/21/18. On 09/20/2019, a System Correction Notice was sent to Cath/EP Lab Managers informing them that the firm will begin the process of upgrading the firmware of affected cardiac ablation systems. The firmware update will eliminate one potential cause of this behavior. Residual risks for other potential causes of RF termination delay remain. Cath/EP Lab Managers were asked to distribute the notice within their respective facilities, continue to post a copy of Appendix A of the June 2018 notice near each affected system, allow the recalling firm to access affected systems for the firmware update, and to complete, sign, and return the acknowledgement form.
- Quantity in commerce
- 793
- Distribution
- U.S.: PA, IL, NY, CA, SC, AR, SD, AZ, KY, FL, TX, MA, MI, OH, LA, VA, NC, WA, DE, IN, TN, NJ, CT, AL, ME, GA, NH, MN, MD, DC, OK, HI, WI, CO, MO, IA, GU, NE, OR, ID, UT, RI, KS, NV, MS, WV Foreign (OUS): Canada, Albania, Austria, Belgium, Bulgaria, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Lithuania, Namibia, Netherlands, Lebanon, Pakistan, Poland, Portugal, Russian Fed., Saudi Arabia, South Africa, Spain, Sudan, Sweden, Switzerland, Turkey, Unit.Arab Emir., Argentina, Australia, Bangladesh, Chile, China, Colombia, Japan, Malaysia, Mexico, Nicaragua, Panama, Philippines, Sri Lanka, Taiwan, Thailand, Vietnam
- Product codes and lots
- All Maestro 4000 Generators used with the optional foot switch.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0202-2019 | Class II | Ongoing | Voluntary: Firm initiated |