Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric...
Luminex Corporation began a recall of Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric bacteria, viruses, and genetic virulence markers from... on . FDA classified it as Class II and gives the reason as "Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2952-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Luminex Corporation
- Product
- Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric bacteria, viruses, and genetic virulence markers from liquid or soft stool preserved in Cary-Blair medium, collected from individuals with signs and symptoms of gastrointestinal infection.
- Product code
- PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
- Reason for recall
- Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023).
- FDA root cause
- Under Investigation by firm
- Action
- The firm notified customers on or about 05/08/2018 via a letter entitled "URGENT: MEDICAL DEVICE RECALL". The Luminex informed the customer reason why they are receiving a recall notice---elevated false positive Yersinia enterocolitica (Yersinia) results when using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023). The letter described the intended use of the test kit and the distribution dates. The letter provided instructions for destroying the product if found in the inventory by a customer at a customer site. Luminex advised the customer to complete and return an enclosed Acknowledgment and Receipt form whether a customer had a recalled product on site or not. Luminex Technical Support are available assist in completing this form. Luminex Customer Experience Team contact information is as follows: 1-877-785-2323 (U.S. and Canada) (24 hours/7days) +1-512-381-4397 (Outside U.S. and Canada) (24 hours/7days) [email protected]
- Quantity in commerce
- 790 Cartridge Carriers containing 3160 Cartridges
- Distribution
- Domestic Distribution: Arizona, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Michigan, Minnesota, Missouri, New Hampshire, Tennessee, Texas, Utah, West Virginia, Wisconsin.
- Product codes and lots
- Catalog number 20-005-023; Cartridge Catalog number 20-006-023; Lot 032718023A; UDI 00857573006041
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2952-2018 | Class II | Terminated | Voluntary: Firm initiated |