510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2597-2018Class II

power cord for: (a) AMIA Automated PD Cycler

Baxter Healthcare Corporation / initiated 2018-07-12 / Terminated

Baxter Healthcare Corporation began a recall of power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T5C8500 (Japanese distribution only) on . FDA classified it as Class II and gives the reason as "There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2597-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T5C8500 (Japanese distribution only)
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
FDA root cause
Under Investigation by firm
Action
Method of Notification: A Safety Alert communication will be sent to affected customers via U.S.P.S., first class mail. Mitigating Instructions: Baxter is asking customers to: 1. Unplug the devices from the wall before unplugging the power cord from the back of the cycler. 2. Confirm the receipt of the letter by completing the enclosed customer reply form and returning it to Baxter by either mail, fax, or scanned email.
Quantity in commerce
5445 units
Distribution
US and Japan
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2597-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2597-2018Class IITerminatedVoluntary: Firm initiated