Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal...
Medtronic Sofamor Danek USA Inc began a recall of Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic... on . FDA classified it as Class II and gives the reason as "This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2603-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.MULTI-AXIAL SCREW FOR 5.5/6.0MM RODS, REF 55840005045
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees, delivered 09 July 2018. The notice stated provided a list of Loaner kits where the screws may also be found and provided the following instructions: " Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via email to [email protected]. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product. Transmission of this Field Safety Notice: 1.) This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 2.) Please transfer this notice to other organizations on which this action has an impact. 3.) Please maintain a copy of this notice in your records."
- Quantity in commerce
- 47 units
- Distribution
- US
- Product codes and lots
- lot 0648247W
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2603-2018 | Class II | Terminated | Voluntary: Firm initiated |