Z-2831-2018Class II
S-SPOT REF 777, 777S Adhere to skin for marking and identifying moles in mammography
Beekley Corporation / initiated 2018-06-13 / Terminated / root cause: manufacturing process
Beekley Corporation began a recall of S-SPOT REF 777, 777S Adhere to skin for marking and identifying moles in mammography on . FDA classified it as Class II and gives the reason as "There is a potential for imaging artifact to mimic calcifications on patient's mammograms". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2831-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Beekley Corporation
- Product
- S-SPOT REF 777, 777S Adhere to skin for marking and identifying moles in mammography.
- Product code
- JAC System, X-Ray, Film Marking, Radiographic
- Reason for recall
- There is a potential for imaging artifact to mimic calcifications on patient's mammograms
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On June 14, 2018, Beekley Medical issued an "URGENT MEDICAL DEVICE RECALL" notice dated June 13, 2018 to customers. The notice described the products, problem and actions to be taken. Customers are advised to do the following: - Inspect stock and quarantine affected product - complete and return the acknowledgement response form via fax to: 1-800-735-1234 or mail. - contact firm if inventory of affected product remains with customers at 1-800-233-5539 between 8:00am - 5:30pm EDT. If you have any issues occurred or for further question, contact Quality Assurance & Regulatory Manager at 1-800-233-5539x434.
- Quantity in commerce
- 3,384,024 pieces in total.
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of; AUSTRALIA, AUSTRIA, BAHRAIN, CANADA, GREAT BRITAIN, ISRAEL, SAUDI ARABIA, SOUTH AFRICA & UNITED ARAB EMIRATES.
- Product codes and lots
- REF 777 - Lot Numbers: 02281802 05301801 04041801 04211801 03301801 05101802 05241801 02191801 03151801 03281801 04281801 02131802 REF 777S - Lot Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2830-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2831-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2832-2018 | Class II | Terminated | Voluntary: Firm initiated |