510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2817-2018Class I

bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS...

bioMerieux, Inc. / initiated 2018-01-26 / Terminated

bioMerieux, Inc. began a recall of bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for... on . FDA classified it as Class I and gives the reason as "Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2817-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit bioMerieux VITEK 2 AST-GP75 Test Kit bioMerieux VITEK 2 AST-GP78 Test Kit Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.
FDA root cause
Under Investigation by firm
Action
bioMerieux sent an Urgent Product Correction Notice letter to their French customers on January 26, 2018. Customer letters in the Americas were sent August 21, 2018. The letters identified the affected product, problem, and actions to be taken. Customers should follow steps outlined in the letter and contact bioMerieux for any additional questions regarding these corrective actions. Note: the actions do not include removal of the product. For any questions or concerns, contact your local bioMerieux representative.
Quantity in commerce
13,961,320 US
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Bermuda, Brazil, Canada, China, Curacao, Costa Rica, Dominican Republic, Ecuador, France, Guatemala, Indonesia, Israel, Hong Kong, South Korea, South Africa, Mexico, Myanmar, Nicaragua, Peru, Philippines, Paraguay, Singapore, Thailand, El Salvador, and Vietnam.
Product codes and lots
VITEK 2 AST-P631, Ref # 414961 UDI 03573026426804 OXSF, multiple Gram Positive Cards all lots of cards with Oxacillin and Cefotoxin   OX, OXSF, multiple Gram Positive Cards  all lots of cards with Oxacillin and Cefotoxin  VITEK 2 AST-GP67 Test Kit, Ref: # 22226 UDI 03573026242060, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP71 Test Kit, Ref: #410750 UDI 03573026355173, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP75 Test Kit, Ref: #415670 UDI 03573026437183, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP78 Test Kit, Ref: #421051 UDI 03573026553982, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2817-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2817-2018Class ITerminatedVoluntary: Firm initiated