Z-2925-2018Class II
Stryker Mako Vizadisc Hip Procedure Tracking Kit
Mako Surgical Corp. / initiated 2018-07-24 / Terminated
Mako Surgical Corporation began a recall of Stryker Mako Vizadisc Hip Procedure Tracking Kit on . FDA classified it as Class II and gives the reason as "Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2925-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mako Surgical Corp.
- Product
- Stryker Mako Vizadisc Hip Procedure Tracking Kit
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera.
- FDA root cause
- Under Investigation by firm
- Action
- On about 07/24/2018, Stryker sent letters to all their consignees via URGENT MEDICAL DEVICE REMOVAL letter. Instructions for risk mitigation included to perform a visual inspection the discs during pre-surgery setup as the defect can be recognized due to apparent discoloration and in the event that a Vizadisc is damaged, risk may be mitigated by testing the Vizadiscs by attaching them to the probes and arrays during pre-surgery setup, as per Stryker's recommendations in "Surgical Technique Guide." This provides an opportunity for the user to show each instrument to the camera after assembly to confirm that all discs are detectable. Additional instructions included to inform all individuals of the recall within the organization, complete and return the enclosed Recall Notification Business Reply Form, and arrange for to return all affected devices to Stryker in Mahwah, NJ. On about 11/27/2018, Stryker initiated an expansion of the recall and notified customers via "UPDATE URGENT MEDICAL DEVICE REMOVAL" letter. Instructions are the same as in the original notification letter.
- Quantity in commerce
- 19,906 total devices
- Distribution
- All devices were distributed in the US in the following states: AK, AL, AR, AZ, CA, CO, CT, FL, IA, ID, KS, KY, LA, MA, MD, MI, MN, MO, NE, NC, NH, NJ, NY, NV, OH, PA, RI, SC, SD, TN. TX, UT, VA, WA, WI.
- Product codes and lots
- Catalog Number 107130. Lot Numbers 17097K, 17124K, 17129K, 17136K, 17143K, 17103H. **EXPANSION** Lot Numbers 17007K, 17013K, 17024K, 17035K, 17054K, 16354H.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2924-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2925-2018 | Class II | Terminated | Voluntary: Firm initiated |