Z-2886-2018Class II

Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2018-07-03 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter... on . FDA classified it as Class II and gives the reason as "One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2886-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 Fr. antimicrobial surface catheters in infants weighing less than 2.0, 2.5, and 3.0 kilograms respectively, have not been evaluated. Therefore, the benefits of these catheters in infants weighing less than the indicated weights should be weighed against any possible risks.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On July 3, 2018, the firm sent out Urgent Medical Device Recall letters to customers. The letter communicated the labeling error. Affected consignees are instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot number listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter.
Quantity in commerce
166
Distribution
Distributors in MO and UT. Medical facilities in AL, CA, IL, MA, NH, NM, NY, TX, VA, and WI.
Product codes and lots
13F17K0202

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2886-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2886-2018Class IITerminatedVoluntary: Firm initiated