Brilliance iCT, 728306
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of... on . FDA classified it as Class II and gives the reason as "Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1144-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 29. 2018, the firm sent Customer Information Letters to affected customers. The letter informed customers that Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. Customers were advised to review the user documentation provided with your system related to Quality Assurance checks. If artifacts are detected during review of the images produced by these checks, and are not removed with the recommended steps identified within your user documentation, please contact your service representative before resuming clinical use.
- Quantity in commerce
- 2
- Distribution
- Distributed only to China and UK.
- Product codes and lots
- 85028 100094
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1142-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1143-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1144-2019 | Class II | Terminated | Voluntary: Firm initiated |