Z-2775-2018Class II

Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD

Degania Silicone, Inc. / initiated 2018-06-21 / Terminated / root cause: manufacturing process

Degania Silicone, Ltd. began a recall of Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley... on . FDA classified it as Class II and gives the reason as "The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2775-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Degania Silicone, Inc.
Product
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Product code
EZL Catheter, Retention Type, Balloon
Reason for recall
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Urgent Field Safety Notice letters were distributed to customers on 6/21/18. The letters instructed customers to perform the following: 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: [email protected] Any affected product should be returned to Degania Facility at Rhode Island: Degania Silicone, Inc. 14 Thurber Blvd, Suite A Smithfield, RI 02917, US Contact name: Jackie Dubon e-mail: [email protected]
Quantity in commerce
500
Distribution
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Product codes and lots
S17003407

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2775-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2754-2018Class IITerminatedVoluntary: Firm initiated
Z-2755-2018Class IITerminatedVoluntary: Firm initiated
Z-2756-2018Class IITerminatedVoluntary: Firm initiated
Z-2757-2018Class IITerminatedVoluntary: Firm initiated
Z-2758-2018Class IITerminatedVoluntary: Firm initiated
Z-2759-2018Class IITerminatedVoluntary: Firm initiated
Z-2760-2018Class IITerminatedVoluntary: Firm initiated
Z-2761-2018Class IITerminatedVoluntary: Firm initiated
Z-2762-2018Class IITerminatedVoluntary: Firm initiated
Z-2763-2018Class IITerminatedVoluntary: Firm initiated
Z-2764-2018Class IITerminatedVoluntary: Firm initiated
Z-2765-2018Class IITerminatedVoluntary: Firm initiated
Z-2766-2018Class IITerminatedVoluntary: Firm initiated
Z-2767-2018Class IITerminatedVoluntary: Firm initiated
Z-2768-2018Class IITerminatedVoluntary: Firm initiated
Z-2769-2018Class IITerminatedVoluntary: Firm initiated
Z-2770-2018Class IITerminatedVoluntary: Firm initiated
Z-2771-2018Class IITerminatedVoluntary: Firm initiated
Z-2772-2018Class IITerminatedVoluntary: Firm initiated
Z-2773-2018Class IITerminatedVoluntary: Firm initiated
Z-2774-2018Class IITerminatedVoluntary: Firm initiated
Z-2775-2018Class IITerminatedVoluntary: Firm initiated
Z-2776-2018Class IITerminatedVoluntary: Firm initiated
Z-2777-2018Class IITerminatedVoluntary: Firm initiated
Z-2778-2018Class IITerminatedVoluntary: Firm initiated
Z-2779-2018Class IITerminatedVoluntary: Firm initiated
Z-2780-2018Class IITerminatedVoluntary: Firm initiated
Z-2781-2018Class IITerminatedVoluntary: Firm initiated
Z-2782-2018Class IITerminatedVoluntary: Firm initiated
Z-2783-2018Class IITerminatedVoluntary: Firm initiated
Z-2784-2018Class IITerminatedVoluntary: Firm initiated
Z-2785-2018Class IITerminatedVoluntary: Firm initiated