Z-2946-2018Class II

CIVCO Needle Guide, REF 676-144, QTY 24, STERILE EO

CIVCO Medical Instruments Co., Inc. / initiated 2018-07-30 / Terminated / root cause: manufacturing process

Civco Medical Instruments Co. Inc. began a recall of CIVCO Needle Guide, REF 676-144, QTY 24, STERILE EO on . FDA classified it as Class II and gives the reason as "The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2946-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CIVCO Medical Instruments Co., Inc.
Product
CIVCO Needle Guide, REF 676-144, QTY 24, STERILE EO
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated communication on 07/30/2018 by email and telephone calls. The firm requested return of the recalled product. Questions may be made to the Account Representative by calling CIVCO Medical Solutions, Multi-Modality Imaging at 800-445-6741 or 319-248-6757 7:00 AM - 5:30 PM CST or by email at [email protected]
Quantity in commerce
107 units
Distribution
United States, Austria, Belgium, Denmark, France, Germany, Poland, Spain, Switzerland and United Kingdom
Product codes and lots
UDI 00841436107327, Lot numbers: A018067, A018067, A021026, A021026, A021226, A021226, A021226, A022447, A022447, A023876, A023876,  A025137, A025137, A025137, A025137, A02560, A025606, A025606, A025606, A026609, A026609, A026609, A026609, A026610, A026610, A026610, A028159, A028159, A028645, A029293, A029293, A029293,  A029383, A032036, A032036, A032036, A032036, A032036, A032036, A033139, A033139, A033139, A033139, A033153, A033153, A035009, A035009, A035009, A035010, A035010, A035010, A035010, A035010,  A035568, A035568, A037150, A037231, A037231, A037231, A037585, A037585, A039417, A039417, A040376, A040376, A040839, A041444, A041444, A041444, A041723, A041723

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2946-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2946-2018Class IITerminatedVoluntary: Firm initiated
Z-2947-2018Class IITerminatedVoluntary: Firm initiated