Stryker Neurovascular began a recall of AXS Infinity LS Long Sheath:((6F/.088in)/ x 70cm GEN-10800-70, (6F/.088in)/ x 80cm GEN-10800-80, (6F/.088in)/ x 90cm GEN-10800-90) Made in Ireland, Sterile EO... on . FDA classified it as Class II and gives the reason as "The AXS Infinity LS Long Sheath Dilator outer diameter (OD) specification was exceeded". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3281-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- AXS Infinity LS Long Sheath:((6F/.088in)/ x 70cm GEN-10800-70, (6F/.088in)/ x 80cm GEN-10800-80, (6F/.088in)/ x 90cm GEN-10800-90) Made in Ireland, Sterile EO. The AXS Infinity LS Long Sheath system contents are: One (1) AXS Infinity LS Long Sheath, One (1) Hemostasis valve with extension tubing, One Long Dilator.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- The AXS Infinity LS Long Sheath Dilator outer diameter (OD) specification was exceeded.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 5/2/2018, "Urgent Medical Device Customer Communication" notifications were mailed via courier. Customers were informed to complete and return a Customer Acknowledgement Form. Customers with questions are encouraged to call (510) 413-2100.
- Quantity in commerce
- 15 units
- Distribution
- Only U.S. Consignees: CO,NJ,NY, OH,PA, and WV.
- Product codes and lots
- Lot No. 20156091 ;UPN: GEN-10800-80-S2 UDI (01)07613327298253 (17)201130 (10) 20156091.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3281-2018 | Class II | Terminated | Voluntary: Firm initiated |