Z-3201-2018Class II

Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour...

Medtronic, Inc. / initiated 2018-05-16 / Terminated / root cause: manufacturing process

Medtronic Inc. began a recall of Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour Plus LINK 2.4 wireless blood glucose monitoring... on . FDA classified it as Class II and gives the reason as "CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3201-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour Plus LINK 2.4 wireless blood glucose monitoring system (glucometer, test strips and control solution) is designed for self-testing by persons with diabetes and the monitoring of glucose concentrations for quantitative glucose measurement in fresh capillary full blood removed from the user's fingertips or palms. It is only for in vitro diagnostic use.
Product code
QBJ Integrated Continuous Glucose Monitoring System, Factory Calibrated
Reason for recall
CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL- CONTOUR"PLUS LINK 2.4 Meter (MMT-1151SK - Batch# MD04M051P) Incorrect Unit of Measure" letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: A. Confirm if the Model (REF) of your meter is MMT-1151SK and the batch (LOT) number is MD04M051P. B. If the REF and LOT match this notice, stop using the CONTOUR"PLUS LINK 2.4 meter immediately. C. Call your local Helpline and let the customer service team know you received a notice stating you may have a CONTOUR"PLUS LINK 2.4 meter with the incorrect unit of measure. D. Return your recalled CONTOUR"PLUS LINK 2.4 meter to Medtronic by following the instructions provided by the Helpline representative. Medtronic will provide a free replacement. If you have any questions, contact the Program Manager, Field Corrective Action at 818-576-4700.
Quantity in commerce
2
Distribution
International Distribution to Slovakia only.
Product codes and lots
Lot# MD04M051P S/N# BG1245623B and S/N# BG1245624B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3201-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3201-2018Class IITerminatedVoluntary: Firm initiated