The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16
Intuitive Surgical, Inc. began a recall of The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16 on . FDA classified it as Class II and gives the reason as "An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2835-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
- Root cause tag
- manufacturing process
- FDA root cause
- Use error
- Action
- Intuitive Surgical notified customers via a "Field Safety Notice Urgent Medical Device Correction ISIFA2018-06-C da Vinci Xi Miscalibrated Endoscope Controller" letter. The letter provided: Reason for Field Action, Risk to Health, Affected Products (description, part number, serial number), Action taken by Intuitive Surgical, Actions to be taken by the Customer/ User, contact information for Further Information & Support, and an Acknowledgement Form.
- Quantity in commerce
- 1
- Distribution
- U.S.
- Product codes and lots
- UDI Number: 00886874114193. Serial Number: SK1703.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2835-2018 | Class II | Ongoing | Voluntary: Firm initiated |