Z-2835-2018Class II

The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16

Intuitive Surgical, Inc. / initiated 2018-05-17 / Open, Classified / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16 on . FDA classified it as Class II and gives the reason as "An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2835-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Intuitive Surgical, Inc.
Product
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Root cause tag
manufacturing process
FDA root cause
Use error
Action
Intuitive Surgical notified customers via a "Field Safety Notice Urgent Medical Device Correction ISIFA2018-06-C da Vinci Xi Miscalibrated Endoscope Controller" letter. The letter provided: Reason for Field Action, Risk to Health, Affected Products (description, part number, serial number), Action taken by Intuitive Surgical, Actions to be taken by the Customer/ User, contact information for Further Information & Support, and an Acknowledgement Form.
Quantity in commerce
1
Distribution
U.S.
Product codes and lots
UDI Number: 00886874114193. Serial Number: SK1703.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2835-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2835-2018Class IIOngoingVoluntary: Firm initiated