Z-2888-2018Class II

CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use...

CIVCO Medical Instruments Co., Inc. / initiated 2018-02-23 / Terminated / root cause: design

Civco Medical Instruments Co. Inc. began a recall of CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product... on . FDA classified it as Class II and gives the reason as "A cover failed leak testing during validation testing of new needle guides from a replacement tool". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2888-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CIVCO Medical Instruments Co., Inc.
Product
CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
A cover failed leak testing during validation testing of new needle guides from a replacement tool.
Root cause tag
design
FDA root cause
Device Design
Action
Telephone Script to be read: CIVCO is calling to notify our customers regarding a recall action being initiated. CIVCO has recently discovered a dimensional issue in our Infiniti bracket which may allow the needle guide to a puncture a cover. You are being notified, as your facility has been identified as having received product from one of the affected manufacturing lots. Advise on Action to be taken by Customer or Distributor: To return Infiniti needle guide, contact your Account Representative at 800-445-6741 or 319-248-6757. Alternately you may also use email at [email protected] to schedule the return of your device. Additional information to obtain: Verification of Email contact to send notification letter and response letter.
Quantity in commerce
177 boxes
Distribution
Worldwide Distribution - US Nationwide in the states of GA, KY, CA, NY , NJ, TX, IL, MA, RI, AZ, and IA and countries of Ukraine, Hong Kong, Italy, Netherlands, Switzerland, United Kingdom, Denmark, Germany, Belgium, Spain, Japan, China, and Canada
Product codes and lots
Item: 639-026 - Biopsy Starter Kit for use with Esaote LA332 and SL332 transducers Lot #'s M206950, M664690, M897110, M119380, M605540, M706720, M570610, M543670, M851530, M983170, M358570. M358570, M427660. M526300, M526300, M734570, M912740, and M664690. Item 698-009: Biopsy Starter Kit for use with Zonare C9-3 transducers M427370, M103030, A010473, M103030, M771190, M427390, M293590, M427370, M105150, M089260, M515370, M293580, M888920, M427370, M220960, M558420, M560740, M909270, M096330, M388690, M558220, M695400, M085550, M917410, M419650, M879530, M964370, M093020, M723180, and A010459 Item: 698-016 Biposy Bracket for use with Zonare C9-3 transducers M567620 and M052860 Item 620-116 Biopsy Starter Kit for use with BK Medical 18L5 and 8870 transducers M080890, M559550, M425050, M056610, M092800, M041630, M186540, M122250, M335740, M043310, M499390, M483250, M109060, M275630, M912150, A023904, M457560, M051670, M091360, M895560, A015235, M202260, M869430, M056610, M027090, M308820, M831280, M547370, M253050, M083160, and M865470

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2888-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2888-2018Class IITerminatedVoluntary: Firm initiated