VITROS Chemistry Products CREA Slides, Product Code 680 2584

Ortho-Clinical Diagnostics, Inc. / initiated 2018-07-11 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product Code 680 2584) - For in vitro diagnostic use only... on . FDA classified it as Class III and gives the reason as "The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3193-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product Code 680 2584) - For in vitro diagnostic use only. VITROS Chemistry Products CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Product code
JFY Enzymatic Method, Creatinine
Reason for recall
The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for this assay. The current low end measuring range is 0.05mg/dL for serum and 1.2mg/dL for urine. Based on the CREA testing results for the LoQ, the revised low end CREA measuring range is 0.15mg/dL for serum and 3.2mg/dL for urine.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 7/11/18, Important Product Correction Notification letters were distributed to customers. The letters instructed customers to perform the following actions: VITROS Cl- Slides and VITROS CREA Slides Instructions for Use will be updated to contain the revised information described in this letter; we will notify you upon availability. In the interim, retain a copy of this customer letter for your records. Calibration Diskette and ADD Data Release Version (DRV) 5992 (and above) will support the revised measuring range for VITROS CREA Slides. We anticipate that DRV 5992 will be available beginning on July 28, 2018 by eConnectivity or will be sent via your normal Cal Disk or ADD mailing. Prior to the availability of the revised IFU, be aware that if a serum or plasma sample contains a Triglyceride concentration of 600 mg/dL or higher, the Chloride result may be positively biased by approximately 2.1 mmol/L. Follow your normal laboratory procedures as you would for other known assay interferences. Note: It is acceptable to continue using VITROS Cl- Slides. Upon availability, install the Calibration Diskette or ADD to implement the revised Measuring Range: For VITROS 4600, 5600 or 5,1 FS Systems: Install ADD DRV 5992 or above, Select All Assay Data option For VITROS 250/350 Systems: Install Calibration Diskette DRV 5992 or above Refer to the enclosure for additional instructions to update the Measuring Range. Update laboratory procedures and your Laboratory Information System (LIS), as necessary. Retain this letter until we send a notification to alert you when the revised IFUs are available. Complete the Confirmation of Receipt form and return by July 25, 2018. Post this notification by each system that processes VITROS Cl- or CREA Slides. Forward this notification if the product was distributed outside of your facility.
Quantity in commerce
627543
Distribution
Worldwide distribution. US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.
Product codes and lots
GEN 7, GEN 10, GEN 12, GEN 13, GEN 15, GEN 16, GEN 18

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3193-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3193-2018Class IIITerminatedVoluntary: Firm initiated