VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997
Ortho Clinical Diagnostics Inc began a recall of VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997 on . FDA classified it as Class II and gives the reason as "The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0061-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997
- Product code
- JLW Radioimmunoassay, Thyroid-Stimulating Hormone
- Reason for recall
- The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 12 July 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products (i.e., VITROS 3600, 5600 and/or ECi/ECiQ Systems) to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue using the affected products. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
- Quantity in commerce
- 68230
- Distribution
- Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
- Product codes and lots
- 5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0058-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0059-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0060-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0061-2019 | Class II | Ongoing | Voluntary: Firm initiated |