Haemonetics OrthoPAT System-Integrated Set
Haemonetics Corporation began a recall of Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs... on . FDA classified it as Class II and gives the reason as "OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2950-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Haemonetics Corporation
- Product
- Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
- Product code
- CAC Apparatus, Autotransfusion
- Reason for recall
- OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Haemonetics, notified customers by an email titled "URGENT: MEDICAL DEVICE RECALL" on July 13, 2018 . The email described the product, problem, and actions to be taken. The customers were instructed to inspect product for defect and return units of the affected lots for replacement with non-affected units; complete the attached customer acknowledgement form via FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO: [email protected]. You can contact your local customer service representative at: North America 1.800.537.2802 or email:[email protected].
- Quantity in commerce
- 432 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot # 1017049 , 1150H-00, 1017072
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2950-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2951-2018 | Class II | Terminated | Voluntary: Firm initiated |