Z-0193-2019Class II

Atellica IM 1600 Analyzer (SMN 11066000). The Atellica" Solution is a multi-component system for in vitro...

Siemens Healthcare Diagnostics Inc. / initiated 2018-06-29 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica IM 1600 Analyzer (SMN 11066000). The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens on . FDA classified it as Class II and gives the reason as "Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0193-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica IM 1600 Analyzer (SMN 11066000). The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued an Urgent medical Device Correction, (UMDC ASW18-02.A.US) and Urgent Field Safety Notification (UFSN ASW18-02.A.OUS) on 6/29/18, via FedEx to the US Customer and via email distribution to countries Outside US (OUS). The field action letters explain the behaviors and informed Customers of action to be taken to prevent the behaviors from occurring. Siemens has released an updated software (SW) version (V), 1.14, which corrects the reported behaviors.
Quantity in commerce
92
Distribution
Nationwide Foreign: Australia Austria Belgium Brazil Canada Croatia Czech Republic Denmark France Germany Greece Italy Japan Malaysia Mexico Netherlands Norway Poland Republic Korea Singapore Spain Sweden Switzerland U.A.E. United Kingdom
Product codes and lots
Software (SW) version 1.13 and lower

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0193-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0189-2019Class IITerminatedVoluntary: Firm initiated
Z-0190-2019Class IITerminatedVoluntary: Firm initiated
Z-0191-2019Class IITerminatedVoluntary: Firm initiated
Z-0192-2019Class IITerminatedVoluntary: Firm initiated
Z-0193-2019Class IITerminatedVoluntary: Firm initiated
Z-0194-2019Class IITerminatedVoluntary: Firm initiated