FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable...
Baxter Healthcare Corporation began a recall of FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable Tip Trimmable Tip***Baxter***." The pouches inside... on . FDA classified it as Class II and gives the reason as "Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2957-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable Tip Trimmable Tip***Baxter***." The pouches inside the cartons are labeled in part, "Floseal SPECIAL APPLICATOR TIPS***Baxter***"."
- Product code
- PMX Absorbable Collagen Hemostatic Agent With Thrombin
- Reason for recall
- Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips.
- Root cause tag
- manufacturing process
- FDA root cause
- Mixed-up of materials/components
- Action
- Baxter Healthcare notified customers of the recall on 05/30/2018 via Urgent Product Recall letter to the Directors of Pharmacy. Instructions included to locate and remove all affected product from inventory, contact Baxter Healthcare Center for Service to arrange for return of affected devices, complete and return the enclosed Baxter customer reply form, and notify customers if further distributed. Adverse reactions to the product can be reported to FDA MedWatch Adverse Event Reporting program and/or Baxter Corporate Product Surveillance at 800-437-5176 between the hours of 8;00 AM and 5:00 PM Central Time, Monday through Friday.
- Quantity in commerce
- 46,908 pouches
- Distribution
- Nationwide distribution to AK, AL, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Australia, Korea, Taiwan, Columbia.
- Product codes and lots
- Product Code: 1502186; Box Lot Numbers: HA170718, HA170818, and HA170917; Pouch Lot Numbers: GR338160 and GR338384
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|