Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2957-2018Class II

FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable...

Baxter Healthcare Corporation / initiated 2018-05-30 / Open, Classified / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable Tip Trimmable Tip***Baxter***." The pouches inside... on . FDA classified it as Class II and gives the reason as "Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2957-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
FLOSEAL Special Applicator Tips in cartons labeled in part, "Floseal SPECIAL APPLICATOR TIPS***6 x Malleable Tip Trimmable Tip***Baxter***." The pouches inside the cartons are labeled in part, "Floseal SPECIAL APPLICATOR TIPS***Baxter***"."
Product code
PMX Absorbable Collagen Hemostatic Agent With Thrombin
Reason for recall
Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips.
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
Baxter Healthcare notified customers of the recall on 05/30/2018 via Urgent Product Recall letter to the Directors of Pharmacy. Instructions included to locate and remove all affected product from inventory, contact Baxter Healthcare Center for Service to arrange for return of affected devices, complete and return the enclosed Baxter customer reply form, and notify customers if further distributed. Adverse reactions to the product can be reported to FDA MedWatch Adverse Event Reporting program and/or Baxter Corporate Product Surveillance at 800-437-5176 between the hours of 8;00 AM and 5:00 PM Central Time, Monday through Friday.
Quantity in commerce
46,908 pouches
Distribution
Nationwide distribution to AK, AL, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Australia, Korea, Taiwan, Columbia.
Product codes and lots
Product Code: 1502186; Box Lot Numbers: HA170718, HA170818, and HA170917; Pouch Lot Numbers: GR338160 and GR338384

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2957-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2957-2018Class IIOngoingVoluntary: Firm initiated