Surgical Handpiece, Model: SGA-E2G Order Code: H184, Nakanishi Inc. The SGS/SGA Surgical Handpieces are...
Nakanishi Inc. began a recall of Surgical Handpiece, Model: SGA-E2G Order Code: H184, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes... on . FDA classified it as Class II and gives the reason as "Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0222-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nakanishi Inc.
- Product
- Surgical Handpiece, Model: SGA-E2G Order Code: H184, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.
- Product code
- KMW Handpiece, Rotary Bone Cutting
- Reason for recall
- Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 7/3/2018, the recalling firm, NSK, initiated sending "URGENT: MEDICAL DEVICE RECALL- CORRECTION" notification letters, via email and mail, by YUSEN Logistics. The recalling firm reports that the initial importers were given the responsibility for sub-recalls to their customers and to extend to the wholesaler, the retailer, and the end user (medical user). Customers were informed to Locate the operation manuals associated with your SGS/SGA Surgical Handpiece; Dispose of the any of the following previous operation manuals associated with those handpieces you may have and replace them with new operation manuals enclosed in this mail; Be sure to reprocess your SGS/SGA handpieces before their next use according to the reprocessing instructions listed in the new operation manual. Complete the enclosed Correction Activity Sheet for Users so we can ensure that you have the most up-to-date reprocessing instructions. Please send the Correction Activity Sheet via e-mail, fax, or mail to: Technical Supervisor Service Division NSK America, Corp., 1800 Global Parkway, Hoffman Estates, IL 60192 US Fax: +1-800-838-9328, e-mail: [email protected]. Public (domestic distributors and their end users) are encouraged to call 1-888-675-1675 Ext. 203 and import-to export distributors are encouraged to call +81-3-5828-7816.
- Quantity in commerce
- 6
- Distribution
- U.S. Distribution to states of: CA, FL, GA, IL, WA.; and internationally to: Canada.
- Product codes and lots
- Previous Operation Manual: OM-H0035E001 Model Name: SGA-E2G Order Code: H184 Serial Nos. BL600026 - BL600031
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0219-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0220-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0221-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0222-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0223-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0224-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0225-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0226-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.