Z-2994-2018Class II
Diagnostica Stago STA Noplastine CI, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin...
Diagnostica Stago, Inc. / initiated 2018-07-24 / Terminated / root cause: design
Diagnostica Stago, Inc. began a recall of Diagnostica Stago STA Noplastine CI, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin time (PT) in plasma. REF: 00666 The STA - Noplastine... on . FDA classified it as Class II and gives the reason as "QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2994-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diagnostica Stago, Inc.
- Product
- Diagnostica Stago STA Noplastine CI, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin time (PT) in plasma. REF: 00666 The STA - Noplastine CI kits provide reagents for the determination of the prothrombin time (PT) in plasma with STA-R, STA Compact and STA Satellite. (In the USA this procedure has been assigned to the moderate complexity category per CLIA 1988 - CDC Analyte Code 4922; CDC Test System Codes 4677 and 4875).
- Product code
- GJS Test, Time, Prothrombin
- Reason for recall
- QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Diagnostica Stago issued notification on July 24, 2018 to US Customers provided the reason for the notification, actions to be taken to mitigate the risk. Action: If you have in your laboratory, any of the lots among those listed in the appendix, we are asking you to: If it is not already done, run a Quality Control test at every change of vial. Return to your Stago affiliate, by fax or e-mail, the enclosed form completed and confirming that you have read this letter. Questions contact Stago Hotline (1-800-725-0607)
- Quantity in commerce
- 5 kits
- Distribution
- Nationwide
- Product codes and lots
- Lot Number/Exp. Date/UDI 251426 2018-11-30 (01)03607450006667(11)161130(17)181130(10)251426(241)00666; 251763 2019-02-28 (01)03607450006667(11)170228(17)190228(10)251763(241)00666
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2993-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2994-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.