Z-0842-2019Class II

Aplicare Povidone Iodine Prep Pads packaged within Portex Pro-Vent Arterial Blood Sampling Kit with Dry...

Abbott Medical / initiated 2018-06-22 / Terminated / root cause: labeling and UDI

Smiths Medical ASD Inc. began a recall of Aplicare Povidone Iodine Prep Pads packaged within Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, models... on . FDA classified it as Class II and gives the reason as "The povidone iodine prep pads packaged with the arterial blood sampling kits will not maintain the iodine potency through its labeled 36-month expiration dating". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0842-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Aplicare Povidone Iodine Prep Pads packaged within Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, models 4589P-1, 4591EG-1, 4591P-1, 4599P-1, 4611LH, 4611P-1, 4620EG-1, 4620P-1, 4640LH, 4640P-1, 4649P-1, 4660EG-1, 4660P-1, 4699LH, 4699P-1, G1325, G1412, G1481, G1509, G1548, G1581, G1586, G1602, G1627, G1632, G1634, G1671, G1675, G1716, G1724, G1773, and G1807. The firm name on the label is Smiths Medical ASD, Inc., Keene, NH.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
The povidone iodine prep pads packaged with the arterial blood sampling kits will not maintain the iodine potency through its labeled 36-month expiration dating.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
The recalling firm issued letters via UPS tracked mail on 6/22/2018 and later again to additional customers on 7/19/2018. An amended letter was issued on 9/20/2018 via UPS tracked mail to include some model numbers that were inadvertently omitted in the previous letter issuance.
Quantity in commerce
Approx. 28,578,939 kits for both products
Distribution
Distribution was nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahamas, Costa Rica, Latvia, and Saudi Arabia.
Product codes and lots
Product with expiration dates between 11/30/2016 and 5/31/2020.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0842-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0842-2019Class IITerminatedVoluntary: Firm initiated
Z-0843-2019Class IITerminatedVoluntary: Firm initiated