Z-2917-2018Class II
MOSAIQ Oncology Information System
Elekta Limited / initiated 2018-08-03 / Terminated
Elekta, Inc. began a recall of MOSAIQ Oncology Information System on . FDA classified it as Class II and gives the reason as "Order Status Remains Approved When Should Indicate Complete". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2917-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- MOSAIQ Oncology Information System
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Order Status Remains Approved When Should Indicate Complete.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- On August 2018 the firm sent letter to all their customer with the following instructions; Recommended User Action: A user with appropriate pharmacy rights can unlock the record. This blanks out the details requiring the user to re-enter the administration details previously entered. To avoid this locking issue, close the MAR window before navigating to the CWS. This document contains important information for the continued safe and proper use of your equipment. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. Advise the appropriate personnel working with this product the content of this letter. Elekta Corrective Actions: The issue is resolved in MOSAIQ Release 2.65 Beta 02 and in release 2.64 SP10. In order to meet regulatory requirements, you are required to complete this the Acknowledgement Form and return it to Elekta immediately upon receipt but no later than 30 days.
- Quantity in commerce
- 185
- Distribution
- AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, WA, WI, WV, and WY Australia, Canada, Ireland, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, and United Kingdom
- Product codes and lots
- Software Build(s) 2.64 SP3 and higher