Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2917-2018Class II

MOSAIQ Oncology Information System

Elekta Limited / initiated 2018-08-03 / Terminated

Elekta, Inc. began a recall of MOSAIQ Oncology Information System on . FDA classified it as Class II and gives the reason as "Order Status Remains Approved When Should Indicate Complete". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2917-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
MOSAIQ Oncology Information System
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Order Status Remains Approved When Should Indicate Complete.
FDA root cause
Unknown/Undetermined by firm
Action
On August 2018 the firm sent letter to all their customer with the following instructions; Recommended User Action: A user with appropriate pharmacy rights can unlock the record. This blanks out the details requiring the user to re-enter the administration details previously entered. To avoid this locking issue, close the MAR window before navigating to the CWS. This document contains important information for the continued safe and proper use of your equipment. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. Advise the appropriate personnel working with this product the content of this letter. Elekta Corrective Actions: The issue is resolved in MOSAIQ Release 2.65 Beta 02 and in release 2.64 SP10. In order to meet regulatory requirements, you are required to complete this the Acknowledgement Form and return it to Elekta immediately upon receipt but no later than 30 days.
Quantity in commerce
185
Distribution
AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, WA, WI, WV, and WY Australia, Canada, Ireland, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, and United Kingdom
Product codes and lots
Software Build(s) 2.64 SP3 and higher

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2917-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2917-2018Class IITerminatedVoluntary: Firm initiated