Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The...
Smiths Medical ASD Inc. began a recall of Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a... on . FDA classified it as Class II and gives the reason as "Several lots of Jelco ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1218-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Several lots of Jelco ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Smiths Medical delivered an "URGENT MEDICAL DEVICE RECALL NOTICE" dated 8/22/2018 and accompanying Recall Response Forms via tracked mail on 08/22/2018 to all consignees who purchased potentially affected product. The notices describe the product, problem and actions to be taken. The consignees were instructed to do the following: 1. Locate affected Jelco ViaValve" safety I.V. catheters in your possession 2. Determine the number of affected devices in your possession and complete the Recall Response Form within 10 days of receipt, returning it to [email protected]. The form must be returned even if you do not have any of the affected product in your possession. Product credit and/or replacement will be processed once the Recall Response Form is received and returned affected product has been verified. 3. All affected product must be returned to Stericycle for processing. Please include a copy of the completed Recall Response Form inside EACH BOX of returned product to facilitate processing. 4. DISTRIBUTORS: If you have distributed potentially affected product to your customers, please immediately notify them of this recall. If you have any questions, contact the Compliance Manager at 763-383-3310 or email to: [email protected].
- Quantity in commerce
- 136,600 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.
- Product codes and lots
- Lot Numbers: 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1218-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY