Z-1218-2019Class II

Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The...

Abbott Medical / initiated 2018-08-22 / Terminated / root cause: manufacturing process

Smiths Medical ASD Inc. began a recall of Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a... on . FDA classified it as Class II and gives the reason as "Several lots of Jelco ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1218-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Several lots of Jelco ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Smiths Medical delivered an "URGENT MEDICAL DEVICE RECALL NOTICE" dated 8/22/2018 and accompanying Recall Response Forms via tracked mail on 08/22/2018 to all consignees who purchased potentially affected product. The notices describe the product, problem and actions to be taken. The consignees were instructed to do the following: 1. Locate affected Jelco ViaValve" safety I.V. catheters in your possession 2. Determine the number of affected devices in your possession and complete the Recall Response Form within 10 days of receipt, returning it to [email protected]. The form must be returned even if you do not have any of the affected product in your possession. Product credit and/or replacement will be processed once the Recall Response Form is received and returned affected product has been verified. 3. All affected product must be returned to Stericycle for processing. Please include a copy of the completed Recall Response Form inside EACH BOX of returned product to facilitate processing. 4. DISTRIBUTORS: If you have distributed potentially affected product to your customers, please immediately notify them of this recall. If you have any questions, contact the Compliance Manager at 763-383-3310 or email to: [email protected].
Quantity in commerce
136,600 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.
Product codes and lots
Lot Numbers: 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1218-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1218-2019Class IITerminatedVoluntary: Firm initiated