ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000
Siemens Medical Solutions USA, Inc began a recall of ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000 on . FDA classified it as Class II and gives the reason as "Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3190-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Medical Device Correction notification letters dated 7/27/18 were distributed to customers. The letters informed the customers of the following planned actions: Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX001/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
- Quantity in commerce
- 429 total
- Distribution
- US nationwide distribution.
- Product codes and lots
- 164013 164011
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3189-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3190-2018 | Class II | Terminated | Voluntary: Firm initiated |