Z-3190-2018Class II

ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000

Siemens Medical Solutions USA, Inc. / initiated 2018-07-29 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000 on . FDA classified it as Class II and gives the reason as "Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3190-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Medical Device Correction notification letters dated 7/27/18 were distributed to customers. The letters informed the customers of the following planned actions: Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX001/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
Quantity in commerce
429 total
Distribution
US nationwide distribution.
Product codes and lots
164013 164011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3190-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3189-2018Class IITerminatedVoluntary: Firm initiated
Z-3190-2018Class IITerminatedVoluntary: Firm initiated