Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED PG UPN: H965440210 Catalog Number: 44-021 Box of 1 Port -... on . FDA classified it as Class II and gives the reason as "BioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium Valved Non-Filled, when it should have been, 8F SL BioFlo Plastic...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0076-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED PG UPN: H965440210 Catalog Number: 44-021 Box of 1 Port - Product Usage - Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- BioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium Valved Non-Filled, when it should have been, 8F SL BioFlo Plastic Valved Non-Filled
- Root cause tag
- sterility
- FDA root cause
- Unknown/Undetermined by firm
- Action
- On July 24, 2018 AngioDynamics initiated this recall action by notifying the affected consignee via telephone call.Per the telephone conversation, the consignee was instructed to segregate the affected device and quarantine the product until further direction from AngioDynamics. A follow-up letter issued dated 7/24/18 email response form to be completed and returned.
- Quantity in commerce
- 1 box
- Distribution
- US Nationwide in the state of IN.
- Product codes and lots
- Lot Number: 5363172