Z-3253-2018Class III

Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35

Medtronic, Inc. / initiated 2018-08-16 / Terminated / root cause: labeling and UDI

Medtronic Vascular began a recall of Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35 on . FDA classified it as Class III and gives the reason as "Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3253-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item, the outer carton has all the correct labeling, and all the other information and configuration details on both the carton and the pouch are correct.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Medtronic initiated consignee communications on 16-Aug-2018 to Risk Managers and Health Care Professionals. US notifications were sent via 2-Day UPS delivery. Consignees are asked to quarantine and return any unused potentially affected product to Medtronic.
Quantity in commerce
250 units
Distribution
CA, IA, NE, TN, TX, VA, WA, Jamaica, New Zealand, Thailand
Product codes and lots
UDI/GTIN: 20643169738717 Lot Number 60068186

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3253-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3253-2018Class IIITerminatedVoluntary: Firm initiated