Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35
Medtronic Vascular began a recall of Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35 on . FDA classified it as Class III and gives the reason as "Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3253-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item, the outer carton has all the correct labeling, and all the other information and configuration details on both the carton and the pouch are correct.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Medtronic initiated consignee communications on 16-Aug-2018 to Risk Managers and Health Care Professionals. US notifications were sent via 2-Day UPS delivery. Consignees are asked to quarantine and return any unused potentially affected product to Medtronic.
- Quantity in commerce
- 250 units
- Distribution
- CA, IA, NE, TN, TX, VA, WA, Jamaica, New Zealand, Thailand
- Product codes and lots
- UDI/GTIN: 20643169738717 Lot Number 60068186
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3253-2018 | Class III | Terminated | Voluntary: Firm initiated |