CyPass ULTRA SYSTEM, CYPASS MICRO-STENT, CYPASS APPLIER, REF CATALOG #: 8065754007. For ophthalmic use
Alcon Research, LTD. began a recall of CyPass ULTRA SYSTEM, CYPASS MICRO-STENT, CYPASS APPLIER, REF CATALOG #: 8065754007. For ophthalmic use on . FDA classified it as Class I and gives the reason as "Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0187-2019
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- CyPass ULTRA SYSTEM, CYPASS MICRO-STENT, CYPASS APPLIER, REF CATALOG #: 8065754007. For ophthalmic use.
- Product code
- OGO Intraocular Pressure Lowering Implant
- Reason for recall
- Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.
- Quantity in commerce
- 38682 units
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
- Product codes and lots
- GTIN: 00380657540075, 10380657540072, 20380657540079 All lot/serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0185-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0186-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0187-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0188-2019 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
