Z-0069-2019Class II

ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular...

Abbott Medical / initiated 2018-08-20 / Completed / root cause: design

ICU Medical, Inc. began a recall of ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile... on . FDA classified it as Class II and gives the reason as "Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use...". As of , the recall status on record is Completed, and MDM tags the root cause as design.

Recall record

Recall number
Z-0069-2019
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Abbott Medical
Product
ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Product code
FMG Stopcock, I.V. Set
Reason for recall
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Root cause tag
design
FDA root cause
Device Design
Action
On 08/20/18, the firm, icumedical, mailed "URGENT MEDICAL DEVICE RECALL" l notices to consignees. The notice described the product, problem and actions to be taken. If product has been further distributed, the firm requested consignees to notify users. Customers were advised to return affected product only if the blood filter is needed. Customers needing response forms and/or return labels were advised to call 1-800-479-0551. For further inquiries call: 1-866-829-9025.
Quantity in commerce
2361 Units
Distribution
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Product codes and lots
lots: 85-540-SJ, 86-189-SL, 86-911-SJ

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0069-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0069-2019Class IICompletedVoluntary: Firm initiated