Z-3242-2018Class II

Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures...

Terumo Cardiovascular Systems Corporation / initiated 2018-08-28 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763 on . FDA classified it as Class II and gives the reason as "Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3242-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Terumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET
Quantity in commerce
47 units
Distribution
Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam
Product codes and lots
Serial Numbers: 0386  1481  1497  1509  1526  1529  1531  1532  1535  1536  1537  1538  1539  1540  1541  1542  1543  1544  1545  1546  1548  1549  1550  1551  1552  1553  1554  1555  1556  1557  1558  1561  1562  1563  1564  1565  1566  1570  1597  1612  1630  1631  1640  1641  1649  1547  1560  1559  1567  Spare Parts:  870509 ASSEMBLY POWER MANAGER POD APS S/N: 03130 , 03308

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3242-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3242-2018Class IITerminatedVoluntary: Firm initiated
Z-3243-2018Class IITerminatedVoluntary: Firm initiated