Z-3243-2018Class II

Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures...

Terumo Cardiovascular Systems Corporation / initiated 2018-08-28 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764 on . FDA classified it as Class II and gives the reason as "Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3243-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Terumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET
Quantity in commerce
10 units
Distribution
Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam
Product codes and lots
Serial Numbers: 1522  1523  1524  1525  1528  1530  1526  1527  1532  1544  Spare Parts:  870509 ASSEMBLY POWER MANAGER POD APS S/N: 03097  03099  03100  03101  03113  03114  03115  03143  03155  03167  03264

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3243-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3242-2018Class IITerminatedVoluntary: Firm initiated
Z-3243-2018Class IITerminatedVoluntary: Firm initiated