Z-0067-2019Class II

Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420

HeartWare, Inc. / initiated 2018-08-28 / Terminated / root cause: design

Heartware, Inc. began a recall of Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for... on . FDA classified it as Class II and gives the reason as "Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0067-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HeartWare, Inc.
Product
Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic began notifying Clinicians in the US on 28-Aug-2018 mainly via 2-day courier, of the identified issues and provided patient management recommendations. Outside the US, Medtronic began notifying Clinicians on 30-Aug-2018 via locally approved methods.
Quantity in commerce
17,275 units
Distribution
Worldwide Distribution
Product codes and lots
Serial number range: CON300175 - CON320540

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0067-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0067-2019Class IITerminatedVoluntary: Firm initiated