Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420
Heartware, Inc. began a recall of Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for... on . FDA classified it as Class II and gives the reason as "Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0067-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- HeartWare, Inc.
- Product
- Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Medtronic began notifying Clinicians in the US on 28-Aug-2018 mainly via 2-day courier, of the identified issues and provided patient management recommendations. Outside the US, Medtronic began notifying Clinicians on 30-Aug-2018 via locally approved methods.
- Quantity in commerce
- 17,275 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Serial number range: CON300175 - CON320540
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0067-2019 | Class II | Terminated | Voluntary: Firm initiated |