Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman...
Beckman Coulter Inc. began a recall of Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes... on . FDA classified it as Class III and gives the reason as "Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Drive software version 4.2 was replaced...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0358-2019
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes the Sorting-Drive 4.2.1 User Kit which includes the Sorting-Drive 4.2.1 Software DVD. . The Sorting-Drive software facilitates the Laboratory Information System (LIS) connection to AutoMate 2500 Family systems. .
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 6/4/2018 Beckman Coulter (BEC) reported that the issue will be corrected via a field modification performed by Field Service. BEC reported that they do not intend to send a customer letter for this issue.US customers can contact the Technical Support Troubleshooting Hotline at (800) 854-3633.
- Quantity in commerce
- 495
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar, Romania, Russian Federation, Saudi Arabia ,South Africa , Spain, Sweden, Taiwan, Turkey , and United Kingdom.
- Product codes and lots
- v4.2.1 (and 4.2, discontinued)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0353-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0354-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0355-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0356-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0357-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0358-2019 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0359-2019 | Class III | Terminated | Voluntary: Firm initiated |