Retractor System Elastic Stays (Part 3314-lG; 3316-lG; 3550-lG)
CooperSurgical, Inc. began a recall of Retractor System Elastic Stays (Part 3314-lG; 3316-lG; 3550-lG) Product Usage: The Retractor System Elastic Stays (Part 3314-lG; 3316-lG; 3550-lG) are a... on . FDA classified it as Class II and gives the reason as "There is a possibility that the seal of the sterile pouch may be compromised, thereby increasing the risk of infection". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3278-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- Retractor System Elastic Stays (Part 3314-lG; 3316-lG; 3550-lG) Product Usage: The Retractor System Elastic Stays (Part 3314-lG; 3316-lG; 3550-lG) are a sterile single-use medical device providing retraction to achieve and maintain optimal visualization throughout a variety of procedures.
- Product code
- GAD Retractor
- Reason for recall
- There is a possibility that the seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Urgent Medical Device Recall Notification letters dated 9/4/18 were distributed to customers. The letters instruct customers to do the following: Please discontinue use of the product with any packaging irregularities, quarantine the product, and complete the attached Acknowledgement and Receipt Form to arrange for either a product replacement or credit to your account through CooperSurgical. Once completed and returned to CooperSurgical, a Customer Service Representative will contact you with a Return Merchandise Authorization (RMA) number and provide instructions for the return of product to CooperSurgical. If replacements are requested, a replacement order will be placed immediately. If you have no affected stock, please complete and return the enclosed Acknowledgement and Receipt Form, in order for us to document receipt of this letter.
- Quantity in commerce
- 18,000 units
- Distribution
- Worldwide Distribution n- US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NY, OH, OR, PA, SC, TX, VA, WA and WV, and the following foreign countries of: Australia, Canada, Colombia, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, Russian Federation, Saudi Arabia, Singapore, Spain, Switzerland, Turkey, and the United Kingdom.
- Product codes and lots
- Lots 244660, 245964, 250426, 251395 , 244667, 244883, 244884, 245603, 245973, 246138, & 248871
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3278-2018 | Class II | Terminated | Voluntary: Firm initiated |