Z-0172-2019Class II
Gibco DPBS, calcium, magnesium, for In-Vitro Diagnostic Use SKU 14040141
Life Technologies Corporation / initiated 2018-08-24 / Terminated / root cause: sterility
Life Technologies, Corp. began a recall of Gibco DPBS, calcium, magnesium, for In-Vitro Diagnostic Use SKU 14040141 on . FDA classified it as Class II and gives the reason as "Leaky bottles due to a defect in the bottle cap compromising the sterility of the product". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0172-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Life Technologies Corporation
- Product
- Gibco DPBS, calcium, magnesium, for In-Vitro Diagnostic Use SKU 14040141
- Product code
- KIT Media And Components, Synthetic Cell And Tissue Culture
- Reason for recall
- Leaky bottles due to a defect in the bottle cap compromising the sterility of the product
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- Thermo Fisherr issued Urgent Medical Device Recall Notification Letter with a customer response sheet sent to consignees via FedEx on 24-Aug-2018. Consignees have been instructed to discard affected product per local and state requirements, or return the affected product to Thermo Fisher Scientific. All consignees are instructed to complete the attached Customer Response Sheet. Thermo Fisher Scientific will issue a no-cost replacement for the recalled product. In addition, consignees are instructed to review results and laboratory reporting associated with use of the affected product for further action. Consignees should notify any personnel who need to be aware of the potential for failure of performance for this product. Questions: contact Technical Services Department at 800-955-6288 Option 2.
- Quantity in commerce
- 240 units
- Distribution
- Nationwide: Foreign: Argentina Australia Canada China Great Britain Hong Kong India Japan Mexico Netherlands New Zealand Singapore South Korea Taiwan
- Product codes and lots
- Lot# 1994520 Exp. date: 30-Jun-2021 (UDI 10190302002691)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0171-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0172-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0173-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0174-2019 | Class II | Terminated | Voluntary: Firm initiated |